An LED to indicate the state of charge of the battery:
A.2.7
Batteries and charger
Precautions to be taken if battery charge is low:
On reception of your THERMYS 150 it can happen that the batteries are not sufficiently charged for optimum operation or even for starting up the device.
It is therefore required to connect the device to the mains (see paragraph A.2.4) and to wait for a few minutes before starting it up (by pressing the On/Off
button).
During normal operation:
When the
symbol is flashing, the battery should be recharged as soon as possible. Connect the charger to power system, the charge indicator (red LED)
on the front panel lights up. Leave the charger switched on for about 3 hours for a complete recharge and disconnect the charger when the charge
indicator
goes off on the front panel.
Precautions to be taken to improve the service life of your batteries:
The battery technology (NiMH: nickel-metal hydride) used in your THERMYS 150 provides greater autonomy, however this technology requires strict
maintenance, requiring discharge cycles to avoid "memory effect". Any nickel-based battery should be fully discharged once a month. If such maintenance is not
performed, a loss of capacity can be seen that can amount to a third of the capacity. Complete restoration then becomes more difficult if such regular
maintenance is ignored.
Warning
ONLY USE THE MAINS ADAPTOR SUPPLIED WITH THE CALIBRATOR
A.2.8
Replacing the battery pack
The battery pack should be changed once a year for optimal autonomy of instrument. To replace the battery pack, contact your dealer.
A.2.9
Stand
The stand gives a good angle of view when the THERMYS 150 is placed on a desk. Unfold the stand on the back of the unit and place the THERMYS 150 on a
desk as shown below.
A.2.10
Strap
The THERMYS 150 is supplied with an untearable strap and two pins to attach the strap to the case.
Before attaching the strap, pass the free end through the fixing loop as in the diagram.
Feed the ends of the strap through the two slots on each side of the case. Insert the two pins into the strap and pull the strap to lock the pins in the case.
.
A
Material
63/107