A.7.1 Resistive Temperature Probes (Temperature) - AOIP THERMYS 150 Mode D'emploi

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

THERMYS 150 G
ENERAL
The MEASUREMENT MENU dialogue box is displayed.
A.7.1
Resistive temperature probes (Temperature)
Display the MEASUREMENT MENU dialogue box:
Select the Rt measurement function, then the appropriate "type of probe" using the function and navigation keys.
Select Auto, 2 -Wire, 3-Wire or 4-Wire connection with the function and navigation keys.
Select display unit.
Confirm with ENTER.
Connection is made according to the number of wires of the resistive probe.
The following probes are available:
α Being the temperature coefficient of the probe.
Wiring diagram is automatically detected by the calibrator. The THERMYS 150 displays an icon showing the connections used (
3 wires or
for 4 wires) to carry out the measurement.
In order not to introduce an error when measuring with 3 wires, the following is recommended:
Measure using conductors of the same length, the same diameter and the same type of metal (a difference of 40 mΩ between two wires
introduces an error of about 0.1°C).
Clean connections to avoid development of spurious electromotive force (emf).
Take care with the connections to avoid the appearance of interfering contact potential differences.
Note:
By changing the display unit, the ohmic value of the connected probe can be measured:
The following ranges are available:
72/107
Sensor
Pt 50(α = 3851)
Pt 100 (α = 3851)
Pt 100 (α = 3916)
Pt 100 (α = 3926)
Pt 200 (α = 3851)
Pt 500 (α = 3851)
Pt 1 000 (α = 3851)
Ni 100 (α = 618)
Ni 120 (α = 672)
Ni 1 000 (α = 618)
Cu 10 (α = 427)
Cu 50 (α = 428)
Probe Type Caption
Pt50
Pt100
Pt100-3916
Pt 100-3926
Pt200
Pt500
Pt1000
Ni100
Ni120
Ni1000
Cu10
Cu50
for 2 wires,