6. Preparing for pumping
Caution
20
l Wash hands thoroughly (at least one
minute) with soap and water before
touching the breast, the clean pump
parts and clean pump set. Dry your
hands with a fresh towel or a single-use
paper towel.
Note
l Use Medela original accessories only.
l Check pump set parts for wear or
damage before use. Replace damaged
or worn parts.
l All parts must be completely dry
before use.
i
Info
l Carry out all the steps with care and
assemble the pump set correctly.
Otherwise a good vacuum may not be
achieved.
6.4 Assembling the pump set
1
Carefully place the membrane into
the connector with the flap into
the opening of the connector.
Make sure that the membrane
forms a seal around the edge of
the connector.
4
Screw the connector onto the
bottle. Choose a bottle size that
suits your needs.