Breathing circuit
Insp./Exp. resistance
1)
at 60 L/min
<4.9 mbar
at 30 L/min
<1 mbar (<1 cmH
at 15 L/min
<0.3 mbar (<0.3 cmH
at 5 L/min
<0.1 mbar (<0.1 cmH
at 2.5 L/min
<0.1 mbar (<0.1 cmH
Compliance at 30 mbar and 60 mbar<1 mL/mbar (<1 mL/cmH
Leakage at 60 mbar
<50 mL/min
Ambient conditions
During storage
Temperature
–20 to 60 °C (–4 to 140 °F)
Relative humidity
5 % to 95 % (non-condensing)
Ambient pressure
500 to 1200 hPa (7.3 psi to 17.4 psi)
During operation
Temperature
5 to 40 °C (41 to 104 °F)
Relative humidity
5 % to 95 % (non-condensing)
Ambient pressure
500 to 1200 hPa (7.3 psi to 17.4 psi)
Classification Medical Device
Class IIa
Europe
UMDNS code
14-238
Universal Medical Device Nomen-
clature System – Nomenclature for
medical devices
Protection class
Type BF
The medical device meets the requirements of the ISO 80601-2-13 stan-
dard.
1)
1 bar = 1 kPa x 100
Order list
Designation
VentStar water trap (P) 180 with Luer Lock
VentStar water trap (P) 180
Infinity ID breathing circuit water trap (P) 180
Additional items according to Dräger accessories
catalog
Instruction for Use VentStar water trap (P) 180 / Infinity ID breathing circuit water trap (P) 180
(<5 cmH
O)
2
O)
2
O)
2
O)
2
O)
2
O)
2
Part number
MP00361
MP00362
MP01341
English / English US
5