amateur radio equipment be closer than 10 m from any part of the BM-02. Interferences can prevent
the BM-02 from monitoring the infant's breathing. Therefore we always recommend you to test the
device functions before use.
o
You should avoid using mattresses made of hard materials (foam polystyrene, etc.) that absorb dis-
turbing vibrations caused by the movement of the ambient air.
o
Bear in mind that the device can only warn you in the case of an emergency but that it does not
prevent the risk of a breathing stoppage! If the baby has any physical problem, it is up to you or your
doctor to provide appropriate help. Do not stay too far from your baby to make sure that you hear the
alarm, should it go off, and that you are able to respond.
o
The manufacturer is responsible for the functionality of the BM-02 provided it is installed and used
in accordance with this manual. The manufacturer waives any responsibility for the proper functiona-
lity of the product in the event that it is mechanically or otherwise damaged or for battery defects.
Moreover, the manufacturer waives any responsibility in the event that the product has been used in
contradiction to this user manual or the service life of the product has been exceeded.
o
The manufacturer urges you not to purchase a used product or to get it from a rental store. Rough
handling may reduce the sensitivity of the sensor, which may have its consequences. Should this be
the case the manufacturer waives all responsibility for the functionality of the product.
The device checks its battery status. When the batteries are running low, necessary replacement is signaled
by the indicator with the battery symbol fl ickering red. Also, the device does not confi rm the activation of the
switch when the batteries are discharged. Turn off the device before replacing the batteries. Remove the
battery compartment cover (see the Installation Instructions – page 4) and take out the original batteries.
Always use only new, branded alkaline AA LR6 batteries (the polarity is clearly marked in the battery
compartment). Once you have replaced the batteries, turn the device on briefl y – its activation must
be confi rmed by a beep. Do not use rechargeable batteries. When fi nished using the device, take the
batteries out.
Besides battery replacement, the device does not require any special maintenance. We recommend
that you occasionally check the sensor pad in the crib to make sure that no moisture precipitates in the
area where the pad touches the mattress. You should turn the pad inside the crib by 180° once in a whi-
le, or turn it over and air it. Clean the pad with a slightly moist cloth. (Do not use aggressive detergents.).
Despite resistance to foreign elements or liquids, their intrusion can damage the device. The product
is supplied with an antibacterial tissue for disinfecting the pad, if necessary. The frequency of cleaning
does not affect the service life of the product. Protect the sensor pad, supply cable and connector from
mechanical damage (shocks, sagging, tensile stress, etc.). If you detect any damage, contact the vendor
or the manufacturer's service department directly (page 11).
1. Why does BM-02 trigger an alarm when the baby is breathing regularly?
Cause: There is not suffi cient contact between the sensor pad and the movement of the baby's body
while the baby is breathing.
Battery replacement
Cleaning and maintenance
Frequently asked questions
7
MFV52124
(859405253BM02ES)