Operation
6
Operation
6.1
Opening and closing the lid
Opening the lid
Closing the lid
6.2
Removing and installing the rotor
Removing the rotor with a
clamping nut
26 / 63
Personnel:
■
Trained user
The centrifuge is switched on
The rotor is stationary.
[STOP/OPEN] button.
Press the
The lid unlocks by means of a motor.
The light on the left side of the
CAUTION
Crushing hazard when closing the lid.
Danger of fingers getting crushed when the closing motor
pulls the lid against the seal.
−
No parts of the operator's body should be in the hazard
zone of the lid when closing the lid.
−
To close the lid, press on the lid from above.
NOTICE
Damage to the device caused by the lid slamming.
−
Close the lid slowly.
−
Do not slam the lid.
When the left side of the [STOP/OPEN] button flashes,
press the [STOP/OPEN] button so that the motorised lid
lock assumes the home position (open).
Personnel:
■
Trained user
Close the lid and press the front edge of the lid down gently.
The lid locks using a motor.
The left side of the
Personnel:
■
Trained user
Rev.: 16 / 11.2023
[STOP/OPEN] button goes out.
[STOP/OPEN] button lights up.