AB1701en
1
4
4
Fig. 21: Transport lock
1
Screws
2
Motor cover
3
Bellows (only for centrifuges with refrigeration)
4
Transport lock
5
Transport lock screws
6
Spacer sleeve
1.
Open the lid.
2.
Unscrew 4 screws (
3.
Remove the motor cover (
4.
For ROTINA 380 R:
Remove the bellows (
5.
Unscrew 3 transport lock screws (
6.
Remove the transport lock screws (
4 ) and store them in a secure place.
port locks (
7.
For ROTINA 380 R:
Insert the bellows (
3 ) over the edge of the rotor vessel. Note the
Push the bellows (
recess for the cable.
8.
Turn over the motor cover (
Rev.: 16 / 11.2023
1
4
1 ).
2 ).
3 ).
5 ).
5 ), spacer sleeves ( 6 ) and trans-
3 ).
2 ) and screw it on.
Commissioning
2
3
23 / 63