Device overview
Fig. 15: [RPM] button
Fig. 16: [START] button
Fig. 17: [T/°C] button
Fig. 18: [TIME] button
Fig. 19: [STOP/OPEN] button
3.6
Original spare parts
3.7
Scope of delivery
18 / 63
■
Querying the integral RCF.
The integral RCF can only be queried if the integral RCF indicator is
enabled.
■
Switch to RCF value.
■
Speed, parameter RPM.
Adjustable from 50 RPM to the maximum rotor speed (N
ments of 10.
■
Switch to RPM value.
■
Start centrifugation runs.
■
Save entries and changes.
■
Go to
'Machine Menu' and open the submenus.
■
Temperature (for centrifuges with cooling)
Adjustable in degrees Celsius (°C) or degrees Fahrenheit (°F).
Parameter T/°C=degrees Celsius (°C). Adjustable from -20°C to +40°C,
in 1°C increments.
Parameter T/°F=degrees Fahrenheit (°F). Adjustable from -4°F to
+104°F, in 1°F increments.
The lowest achievable temperature is rotor dependent.
■
Temperature (for centrifuges with heating)
Enable or disable heating, parameter Heater.
■
Scroll backward in the menus (the button is blank for centrifuges without
cooling).
■
Runtime, parameter t/hms.
h: Hours. from 1 h to 99 h, in 1 hour increments.
m: Minutes. from 1 min to 59 min, in 1 minute increments.
s: Seconds. from 1 s to 59 s, in 1 second increments.
■
Continuous operation "¥"
■
Set the start of the runtime count.
■
End the centrifugation run.
The rotor coasts to a stop with the preselected ramp-down parameter.
■
Pressing the button twice triggers the quick stop function.
■
Unlock the lid.
■
Exit parameter input and the menus.
Use only original spare parts from the manufacturer and approved accesso-
ries.
The following accessories are supplied with the centrifuge:
■
1 grease for the trunnions
■
1 Hex key (5 mm x 170)
■
1 right-angled hex key (2.5 mm)
■
1 6-lobe (Torx) pin key wrench, short (T20 SG)
Rev.: 16 / 11.2023
), in incre-
max
AB1701en