7.6.8
Audible signal
7.6.8.1
General
7.6.8.2
Enabling or disabling an audible signal
AB1701en
[PROG] button for 8 seconds.
1.
Hold down the
'***Machine Menu***' is displayed after 8 seconds.
[PROG] button until '-> Change PIN' is displayed.
2.
Press the
[START] button.
3.
Press the
'old PIN = ---- <START>' is displayed.
��
[PROG] button.
4.
Press the
'Get HELP # no' is displayed.
��
The previous PIN becomes invalid after retrieving the Help number.
[Rotary knob] to set 'yes' .
5.
Use
[START] button.
6.
Press the
'Are you sure ? no' is displayed.
��
[Rotary knob] to set 'yes' .
7.
Use
[START] button.
8.
Press the
'HELP # = 5487' is displayed.
��
Note down this Help number and use it to request the required
PIN. Setting a new PIN using the PIN received
The audible signal sounds:
after a problem occurs in the 2 s interval.
after completion of the centrifugation run and rotor standstill in the 30 s
interval.
Opening the lid or pressing any button stops the audible signal.
The rotor is stationary.
1.
Press and hold the
'***Machine Menu***' is displayed after 8 seconds.
��
[PROG] button repeatedly until '-> Settings' is displayed.
2.
Press the
[START] button.
3.
Press the
'SOUND / BELL = on' or 'SOUND / BELL = off' is displayed.
��
'SOUND / BELL' : Signal after completion of the centrifugation run
[Rotary knob] to set 'off' or 'on' .
4.
Use
off = audible signal disabled
on = audible signal enabled
5.
Press the
[PROG] button.
'SOUND / BELL error = on' or 'SOUND / BELL error = off' is
��
displayed.
'SOUND / BELL error' : Signal after the occurrence of a fault
[Rotary knob] to set 'off' or 'on' .
6.
Use
off = audible signal disabled
on = audible signal enabled