Disposal
Fig. 25: Emergency release
1
Hole
9.4
Switching on the automatic circuit breaker (for types 1701-01 and 1706-01
only)
10
Disposal
10.1
General instructions
60 / 63
1.
Look through the window in the lid to ensure that the rotor is sta-
tionary.
2.
Insert the hex key horizontally into the hole (
until the lid opens.
1
3.
Remove the hex key from the hole (
4.
Check whether the left side of the
power is restored.
When the left side of the
[STOP/OPEN] button so that the motorised lid lock assumes the
home position (open) again.
Personnel:
■
Trained user
Fig. 26: Automatic circuit breaker
1
Plastic pin
The mains switch is in switch position
The centrifuge is disconnected from the mains.
1.
Press the plastic pin (
2.
Reconnect the device to the mains.
The device can be disposed of via the manufacturer.
A Return Material Authorisation (RMA) form must always be
requested for a return.
If necessary, contact the Technical Service Department of
the manufacturer:
−
Andreas Hettich GmbH & Co. KG
−
Föhrenstraße 12
−
78532 Tuttlingen, Germany
−
Phone: +49 7461 705 1400
−
E-Mail: service@hettichlab.com