Caution: Disassembly of certain sub-assemblies can affect operation and safety, for example:
- when intervening on the electrical power supply,
- when intervening on the suspension arms and balancing system,
- when intervening on the optical system of the copulas which are equipped with filters intended to eliminate
radiations not visible by the patient. the unit should not be operated, under any circumstances, without these
filters.
For this type of intervention, contact the MAQUET SA-approved product support service.
AnOMALy
q Cupola does not come
on though green power
supply indicator light is lit.
q Light volume variation
control does not operate.
q All cupolas not lit
q Power supply does not
come on.
q 6000 or 8000 cupola goes
off when handled.
q Unit goes off completely
during an operation
q Low bulb service life
q Sterilizable handle does
not engage correctly on its
support.
46
TROUbLESHOOTInG
PRObAbLE CAUSE
1 - Bulb failure or missing bulb.
2 - Other cause
1 - Power supply indicator light is
red: mains system has failed and
unit is operating on hospital 24V
backup system.
2 - Power cut.
3 - Other cause.
- Each cupola has its own
individual control.
1 - Mains power failure
2 - Other cause
1 - Protection device trips when
cupola exceeds vertical
position.
2 - Other cause
- Power supply failure or
abnormal cupola position.
- Bulbs do not comply or
overvoltage condition is
present.
1 - Average service life of
sterilizable handle: 100
sterilization cycles at 134° ±
4°C for 20 minutes.
2 - Average service life
exceeded / Handle is
deformed.
0113103
CORRECTIVE ACTIOn
Switch off power supply and replace bulb(s).
Call engineering department.
1- Wait for return to normal power system or startup of
hospital power generator.
2- Check if another equipement is connected.
3- Call engineering department.
Check that all the power supplies
are properly switched on
(green indicator light).
Check if another piece of equipment is operating on
the same electrical power system.
Call engineering department.
Return cupola to normal utilization position (see cupola
description, page 15).
Call engineering department.
Check for presence of the main power system on the
control unit:
- green indicator light should be lit. If not lit, call
engineering department.
- if lit, check position of cupola which should not go
past the vertical position.
1) Ensure you are using bulbs recommended by
MAQUET SA. Do not use bulbs used with surgical
lights of previous generation (PRISMATIC) which
are not compatible for power and size.
2) Call engineering department to verify the voltage of
batteries (24 Volts) of emergency supply .
3) Call engineering department to verify AC voltage
at bulb terminals.
(
)
V
RMS
23 V
AC+DC
EFF
Check operation of handle lock mechanism (click) and
overall handle.
Call engineering department.