General information
The device may only be operated by authorized and instructed personnel. Training and qualifi-
cation of the operating personnel as well as ensuring that the device is handled responsibly are
the sole responsibility of the operator!
Transportation
During transportation, avoid severe shocks and mechanical stresses that can cause damage to
the device.
Keep the original packaging in a dry and protected place for later use.
Storage
Always store the device in its original packaging. To protect against damage and excessive
material aging, store the device in an environment that is as dry, temperature-stable and dust-
free as possible.
Acclimatization
After each transportation and after storage under critical climatic conditions (e.g. high
temperature difference between inside and outside), allow the device to acclimatize at room
temperature for at least two hours to prevent possible damage due to condensation before
commissioning it in the place of use. If necessary, extend the acclimatization phase if the
temperature differences are very high.
Make all supply connections (power supply, tubing) only after the device has been acclimatized!
Permissible ambient conditions
The device is designed for indoor use only. Permissible ambient conditions for operation:
•
5 °C – 31 °C at up to 80 % rel. humidity
•
32 °C – 40 °C at up to 50 % rel. humidity (decreasing linearly)
•
Installation altitude: 2,000 m asl
When used in corrosive atmospheres, the service life of the device may be reduced depending
on the concentration, duration and frequency of exposure.
The device IS NOT suitable for outdoor use!
The device IS NOT suitable for use in potentially explosive areas!