Commissioning
Start the calibration process in the device software:
Open the main menu using the
start window:
Settings
Open the
„Main menu" on page 105).
Follow the instructions on the display.
OR
Calibration on high-polar solvents and mixtures:
Make sure the sensor is in the flask.
Fill the evaporation flask with at least 500 ml of the product.
Slide the sensor into the stopped flask until the sensor head is just above the fluid level or
– for example, in the case of mixtures with water – until the sensor head touches the fluid
level slightly.
Start the calibration process in the device software:
Open the main menu using the
start window:
Settings
Open the
„Main menu" on page 105).
Follow the instructions on the display.
PRACTICAL TIP
When processing solvents with a tendency toward boiling retardation
(e.g. ethanol), it is recommended to calibrate the sensor slightly less
sensitively.
This prevents the boiling retardation from being detected as a fluid and
the fluid level in the rotary flask from decreasing excessively before
adding new medium.
The sensor should only respond to liquids, never to boiling retardation or
foam!