Operation
Procedure
Start process
➜
Touch the
–
The vacuum pump is switched on, the pressure in the evaporation flask is lowered to
process pressure.
–
In the footer, the
Fill educt
➜
Touch the
By briefly tapping the
injected into the evaporation flask.
Touch the
button progress) to fill the evaporation flask continuously. The filling has
to be stopped manually!
Fill educt
➜
Touch the
Stop process
➜
Touch the
When the process is ended, the
on the control panel.
Drain distillate
➜
Touch the
Drain distillate
➜
Touch the
The evaporation flask must be drained by hand.
Please observe the relevant information and instructions in the operating
instructions for the rotary evaporator!
button.
Fill educt
Stop process
and
button to start the automatic filling of the evaporation flask.
Fill educt
Fill educt
button for approximately 2 seconds (observe
button again to stop filling.
button to stop the process (end process).
button → Distillate is drained (default: approx. 30 s).
button → Drainage of the distillate is stopped.
buttons are displayed
button, a small amount can be
Drain distillate
140
button appears