Troubleshooting
Troubleshooting
Troubleshooting
Error message
Cable breakage or sensor
defect
Short circuit
Values out of range
Compressor has been
stopped
Pressure sensor failed!
Error due to barometer!
Temperature sensor failed!
Emergency stop has been
triggered!
Educt container S1 is
empty!
Overflow sensor S8 to the
distillate container has
triggered!
Overflow sensor S9 to the
residue container has
triggered!
S3 cut out sensor has
triggered!
Failure message from
recirculation chiller
Invalid sensor plugged in!
Invalid valve plugged in!
Communication error!
Operation mode cannot be
started!
Possible remedy
Check the cable of the displayed component.
Check the cable connections of the displayed
component.
Correct the values of the displayed parameter.
Excessive pressure from compressor, reduce
pressure.
Check pressure sensor for function, system
check.
Check the function of the barometer.
Check temperature sensor for function.
Eliminate the cause of the emergency stop from
the external emergency stop switch.
Refill educt.
Drain the distillate container.
Drain the residue container.
Eliminate the cause of the emergency stop on
the rotary evaporator.
Observe the indicator on the recirculation chiller.
Check the coding of the sensor, check that the
sensor is in perfect condition.
Check the coding of the valve, check that the
valve is in good condition.
Check the cable connection and connectors of
the displayed components.
Open the error list, remove the indicated error
cause(s).