15.4. RF EXPOSURE INFORMATION:
The RS-00500C model meets the applicable limits for radio frequency (RF) exposure, as
determined by independent tests for the Specifi c Absorption Rate (SAR).
Specifi c Absorption Rate (SAR) refers to the rate at which the body absorbs RF energy.
The SAR limit is 1.6 watts for kilogram in countries that set the limit averaged over 1 gram
of tissue and 2.0 watts per kilogram in countries that set the limit averaged over 10 grams
of tissue.
For RS-00500C, with serial numbers beginning ZC, HY, DC, Tthe highest SAR values
EN
measured in accordance with ICNIRP guidelines are:
Body SAR at 1g: 1.50 W/Kg
Body SAR at 10g: 1.70 W/Kg
For RS-00500C, with serial numbers beginning 4C, 4Y, the highest SAR values measured
in accordance with ICNIRP guidelines for this device with a 4/5mm separation as pro-
vided by the system enclosure are:
Body SAR at 1g: 0.742 W/Kg
Body SAR at 10g: 0.834 W/Kg
During use the actual SAR values for this device are usually well below the values stated
above, because for purposes of system e ciency operating power is reduced from full
when not needed and the lower the power output the lower its SAR value.
43/370