15.1.1. Guidance and manufacturer's declaration: electromagnetic emissions
The Zen-O™ Portable Oxygen Concentrator is intended for use in the electromagnetic en-
vironment specifi ed below. The customer or the user of the concentrator should ensure
that it is used in such an environment.
EMISSION TEST
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fl uctuations/
fl icker emissions
IEC 61000-3-3
15.1.2. Guidance and Manufacturer's Declaration: Electromagnetic Immunity
Zen-O™ is intended for use in the electromagnetic environment specifi ed below. The
customer or the user of the concentrator should ensure that it is used in such an environ-
ment.
IMMUNITY
IEC 60601 TEST
TEST
Electrostatic
± 6kV contact
discharge
± 8kV air
(ESD)
IEC 61000-4-2
Electric fast
± 2kV for power
transient/burst
supply lines
IEC 61000-4-4
± 1kV for input/
output lines
COMPLI
ELECTROMAGNETIC ENVIRONMENT/GUIDANCE
ANCE
Group 1
The Zen-O™ Portable Oxygen Concentrator uses RF
energy only for its internal function.
Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic
equipment.
Class B
The concentrator is suitable for use in all establish-
ments, including domestic establishments and those
directly connected to the public low-voltage power
Class A
supply network that supplies buildings used for dome-
stic purposes.
Complies
COMPLIANCE
LEVEL
LEVEL
± 15kV contact
± 8kV air
± 2kV for power
supply lines
± 1kV for input/
output lines
37/370
ELECTROMAGNETIC
ENVIRONMENT/GUIDANCE
Floors should be wood, concrete, or
ceramic tile.
If fl oors are covered with synthetic
material, the relative humidity should
be at least 30%.
Mains power quality should be that
of a typical commercial or hospital
environment.