Indications
CHEK (Figure 3-11)
Pump will sound a dual tone beep once and CHEK, RATE,
DOSE, and FEED INTERVAL will all blink on the display.
Why: This alarm sounds when the RATE, DOSE,
and FEED INTERVAL are not compatible. As
discussed in the Interval Feeding Example
(Page 13), pump divides the dose by the rate
to determine the amount of time required for the dose to be delivered.
Based on the RATE and DOSE combination, when the FEED INTERVAL is
being programmed, the first value available will be the first value that is
compatible with the RATE and DOSE settings. Should the RATE or DOSE
be changed after the FEED INTERVAL has been programmed, and the
combination requires the delivery time to be longer than the amount of
time programmed as the FEED INTERVAL, user will receive this alarm.
For example, if rate is 100 ml/hr, and dose
is 200 ml, then the pump will take 2 hours
to deliver the dose. The first available FEED
INTERVAL will be 2.00 hr. However, if the dose
is changed to 300 ml, the dose will now take
3 hours to deliver and therefore 2.00 hr is an
impossible FEED INTERVAL. If the
key is pressed before the FEED INTERVAL has
been changed (the next available value being
3.00), the pump will give the CHEK alarm.
To correct the problem: Press the
RUN/PAUSE
mode, and reprogram the settings.
TO STOP LET GO (Figure 3-12)
When pressing and holding the
hold to prime, alarm will sound once, pump will
begin pumping at an approximate rate of 700
ml/ hr and display will read TO STOP LET GO.
To stop priming: Release the
key to put pump in pause
PRIME
PRIME
3. ALARMS, MESSAGES, AND INDICATIONS
RUN/PAUSE
key during
key.
33
Figure 3-11 CHEK Indication
Figure 3-12 Manual
Priming Indication