Charging The Battery; Figure 2-28 Fuel Gauge - Moog infinity Manuel De L'opérateur

Masquer les pouces Voir aussi pour infinity:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

2. DIRECTIONS FOR USE
The Battery

Charging the Battery

To charge the battery, insert plug from AC
Adapter/Charger into port on the left side
of pump. Plug charger into a wall outlet.
Pump will charge whether it is
turned off or running.
When charger is plugged in and charging, the
plug symbol will appear in the display and the
fuel gauge will display four bars scrolling from
E to F. This pattern continuously repeats while
pump is charging (Figures 2-28a through 2-28e).
When bars stop scrolling pump is done charging.
To check the status of battery during
charging, disconnect charger from
pump and turn on pump.
NOTE: If pump shuts down due to low
battery, charging less than 10 minutes may
result in an ER99 alarm. Fully charge battery
if depleted (see page 29 to clear alarm).
Infinity
Enteral Feeding Pump Operator's Manual
®
24
a
b
c
d
e

Figure 2-28 Fuel Gauge