Routine Maintenance; Disposal Information - Moog infinity Manuel De L'opérateur

Masquer les pouces Voir aussi pour infinity:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

9. ADDITIONAL TECHNICAL INFORMATION
The Infinity pump is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The
customer or the user of the Infinity pump can help prevent electromagnetic interference by maintaining a minimum distance between
portable and mobile RF communications equipment (transmitters) and the Infinity pump as recommended below, according to the
maximum output power of the communications equipment.
Rated maximum output
power of transmitter
W
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be
estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.

Routine Maintenance

There is no routine calibration or adjustment procedure required for
Infinity.
There are no user-serviceable parts. All service requests
should be referred to Moog technical service.

Disposal Information

At the end of their service lifetime, the pump and its electrical accessories (AC Adapter
Charger and Data Download Accessory Cable) should be disposed of according to
local standards and regulations governing the disposal of electronic waste (e-waste).
Other accessories can be disposed of or recycled as standard non-hazardous waste.
Start Up Graph and Trumpet Graph
Available upon request
Infinity
Enteral Feeding Pump Operator's Manual
®
Recommended separation distances between portable and mobile
RF communications equipment and the Infinity pump
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
d=1.2√P
0.12
0.38
1.2
3.8
12
m
80 MHz to 800 MHz
d=1.2√P
0.12
0.38
1.2
3.8
12
48
800 MHz to 2.7 GHz
d=2.3√P
0.23
0.73
2.3
7.3
23