LIST OF WARNINGS AND CAUTIONS

Be sure to carefully read the following warnings and cautions. Failure to
review all the warning and cautions can cause harm to the operator or
patient. Healthcare personnel who provide training to lay users should be
sure to review all the warning and cautions with the lay user. If there is a
change in the performance of the pump, contact customer service.
WARNINGS
Use ONLY feeding solutions prescribed by the responsible
4
physician, registered dietitian, registered nurse, or other licensed
practitioner. Failure to do so may result in harm to the patient.
Use ONLY Infinity brand disposable sets to ensure proper fluid
4
delivery. Others will not deliver the correct dose, may allow dangerous
free-flow conditions, and may generate hazardous fluid pressures
which may activate occlusion alarms at unpredictable pressures.
Low flow rates, combined with high dose settings may exceed the life
4
of the disposable set. The disposable bag set must be replaced every
24 hours and the spike sets must be replaced every 48 hours (max1.5 L)
to maintain delivery accuracy, allow proper air and occlusion sensing,
and prevent growth of bacteria.DO NOT PROGRAM RATE AND DOSE
COMBINATIONS WHICH EXCEED A 24 HOUR FEED REGIMEN FOR THE
DISPOSABLE BAG SETS OR A 48 HOUR FEED REGIMEN FOR THE SPIKE SETS.
The battery capacity is an approximation. If you are unsure that
4
enough capacity remains for your intended use, recharge it. Failure
to do so may result in delayed or under-delivered therapy.
To avoid electrical shock, never clean pump with charger
4
plugged into an outlet or when pump is on.
To avoid electrical shock, make sure the Infinity AC Adapter/Charger
4
is completely dry before plugging into an electrical outlet.
Do not use Infinity Enteral Feeding Pump for delivery of
4
non-enteral solutions. Serious injury may result.
Proper operation of pump requires door to be closed and latched.
4
Make sure door is closed and latched when motor is running.
If an error occurs (ER01 - ER99 will appear in display), check all settings
4
and volume counters and verify before starting a feeding. Failure
to do so may result in delayed or under-delivered therapy.
Infinity
Enteral Feeding Pump Operator's Manual
®