No modification of this equipment is allowed including the pump,
4
disposable sets, and accessories. Modification can result in delayed
therapy, over or under delivery, electrocution, injury, or death.
Device contains small parts. Dispose of or store connector cap when
4
not in use to avoid choking hazard for children under 3 years.
Avoid leaving tubing and cords where infants, children, or those deemed at high
4
risk for medical line entanglement, can become entangled. Be aware that if these
items become wrapped around a patient's neck, there is an increased risk of
strangulation or death. Caregivers of patients who are at risk for entanglement
should discuss with their health care provider how to properly manage their
lines and properly monitor patients based on their needs while in use.
CAUTIONS
Federal law (U.S.A.) restricts this device to sale by or on the order of a physician,
4
registered dietitian, registered nurse, or other licensed practitioner.
Dispose of the Infinity Pump, Infinity Orange Pump and
4
Disposable Sets properly, as required by local law
If any leaks are detected in the disposable set, stop pump operation
4
and disconnect set from patient and replace with a new disposable
set. Failure to do so may result in under-delivered therapy.
Care should be used when manually priming delivery set
4
to ensure cassette is not damaged by excess force.
Do not overload carry packs with personal items. Pump function may be
4
affected by kinked or pinched tubing and/or unintentional button presses.
The Infinity uses an optical air-in-line sensor to detect air in the tubing
4
as it passes through the pump, causing the pump to alarm if feeding is
inadvertently interrupted or when feeding is complete. It is important
to keep this sensor free of debris including spilled fluids or food that may
adhere to the sensor and prevent proper detection of air in the tubing.
The clinician and caregiver should be trained regarding the care, use,
and maintenance of the Infinity pump and accessories, including
proper cleaning of the area around the optical air-in-line sensor.
Clinicians, care providers, and end users should be aware of the following
considerations whenever using or teaching others to use the pump:
5
© Copyright 2022 Moog, Inc.