2. DIRECTIONS FOR USE
Recommendation for First Use:
Since battery may not be fully charged when pump is first received, it
is recommended that battery be charged for 6 hours prior to operating
on battery power (see page 24 for additional information).
Priming and Loading the Disposable Set
Step 1:
4 If you are using the Infinity Safety
Screw Spike Set with a vented
bottle or pre-filled bag
Remove protective cover from tip, then
insert tip into formula exit port of container.
If using the Infinity Safety Screw Spike Set,
secure the tip by tightly screwing threaded
components together. (Figure 2-1).
4 If you are using the 500ml or
1200ml disposable set:
Hold bag upright and pour in feeding
solution (Figure 2-2). Close cap securely.
NOTE: Blenderized or aggressively mixed
solutions may have foam. If using this type
of solution, allow it to sit for 10 to 15 minutes
before pouring into bag. This will reduce the
chance of an alarm due to air in the tubing.
Step 2:
Remove protective cap from ENFit
connector (Figure 2-3).
NOTE: If set is to be used with a carry pack, all
air must be removed from bag and tubing.
Continue to step 3 for instructions on
removing air. If set is to be hung above pump,
i.e. on an IV pole, you may skip to step 4.
13
2. DIRECTIONS FOR USE
Figure 2-1 Screw Tip
onto Container
Figure 2-2 Fill Bag
Figure 2-3 Remove Protective
Cap from ENFit Connector