NO FOOD Alarm
If re-priming of the pump delivery set is required, it is advisable to disconnect
the enteral adapter from the patient's feeding tube while re-priming.
Is bag of pump
delivery set empty?
NO
Is air present in
the tubing?
NO
Is formula blenderized
or aggressively mixed?
NO
Are air bubbles trapped
inside the cassette?
YES
Refill bag, re-prime delivery set and resume feeding.
Has air been
removed from
YES
bag of pump
delivery set?
Let formula sit for 10-15 minutes before preparing
pump delivery set for feeding. If foam is present in
YES
tubing, flush by priming pump delivery set either
manually or with the pump and resume feeding.
Remove tubing from pump and manually prime tubing
by turning bag upside down, gently squeezing lower
YES
half of bag while pinching the teal colored tubing just
below the drop symbol. Hold this position until all air
has passed through the teal segment of the tubing.
If alarm continues, contact Moog Customer
NO
Service at 800.970.2337 for assistance.
35
4. TROUBLESHOOTING GUIDE
Manually prime tubing by
turning bag upside down,
gently squeezing lower half
of bag while pinching the teal
NO
colored tubing just below the
" " drop symbol. Hold this
position until all air is removed
from bag and tubing.
Press and hold the
key to move the air in the
tubing past pump delivery set
cassette and resume feeding.
YES
If your pump is equipped with
an auto prime option, refer
to page 26 for details.