9. ADDITIONAL TECHNICAL INFORMATION
The Infinity is intended for use in the electromagnetic environment specified below. The customer or the user of the Infinity should
Emissions Test
RF Emissions
CISPR 11
RF Emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/Flicker
emissions
IEC 61000-3-3
The Infinity pump is intended for use in the electromagnetic environment specified below. The customer or the user of the Infinity
pump should assure that it is used in such an environment.
Immunity test
Electrostatic discharge
(ESD)
IEC 61000-4-2
Electrostatic fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
IEC 61000-4-11
Power frequency (50/60
Hz) magnetic field
IEC 61000-4-8
NOTE U
is the a.c. mains voltage prior to application of the test level.
T
Infinity
Enteral Feeding Pump Operator's Manual
®
Guidance and manufacturer's declaration – electromagnetic emissions
assure that it is used in such an environment.
Compliance
Group 1
Class B
Class B
Complies
Guidance and manufacturer's declaration – electromagnetic immunity
IEC 60601 test level
± 8 kV contact
± 15 kV air
± 2 kV for power supply lines
± 1 kV for input/output lines
100 kHz PRR
2kV
(0, 90, 180, & 270° phase)
<0 % U
(>95 % dip in U
) for
T
T
0.5 cycle
40 % U
(60 % dip in U
) for
T
T
5 cycles
70 % U
(30 % dip in U
) for
T
T
25/30 cycles
<0 % U
(>95 % dip in U
) for
T
T
250/300 cycles
30 A/m
Electromagnetic environment - guidance
The Infinity uses RF energy only for its internal function. Therefore, its
RF emissions are very low and are not likely to cause any interference in
nearby electronic equipment.
The Infinity is suitable for use in all establishments, including domestic
establishments and those directly connected to the public low-voltage
power supply network that supplies buildings used for domestic
purposes.
Compliance level
± 8 kV contact
± 15 kV air
± 2 kV for power supply lines
± 1 kV for input/output lines
100 kHz PRR
2kV
(0, 90, 180, & 270° phase)
<0 % U
(>95 % dip in U
) for
T
T
0.5 cycle
40 % U
(60 % dip in U
) for
T
T
5 cycles
70 % U
(30 % dip in U
) for
T
T
25/30 cycles
<0 % U
(>95 % dip in U
) for
T
T
250/300 cycles
Class 2 (30 A/m)
46
Electromagnetic
environment - guidance
Floors should be wood, concrete
or ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment. The Infinity pump
allows continued operation during
power mains interruptions via the
internal battery.
Power frequency magnetic fields
should be at levels characteristic
of a typical location in the typical
commercial or hospital environment.