2. DIRECTIONS FOR USE
Feeding Completion:
When the dose has been completely delivered,
pump will stop running and display will read NEXT
DOSE then give the number of hours and minutes
until feeding will be repeated (Figure 2-26). Display
will show the run symbol flashing without the arcs
rotating around it. Pump will repeat this cycle until
disposable set is empty or until it is stopped by user.
NOTE: When a feed interval has been programmed,
pump will not alarm DOSE DONE. Pump will
cycle continuously as long as feeding solution
remains in the bag. When empty, pump
will alarm NO FOOD or NO FLOW IN.
ON/OFF
To stop feeding, press the
key for 1.5
Figure 2-26 Time Until
RUN/PAUSE
seconds to turn pump off or press the
Next Feeding
key to put pump in pause mode. Adjustments
to settings can be made at this time.
New Feeding Cycle:
Turn pump on. Previous settings for rate, dose, and feed interval
will be saved in memory. If no changes are required, verify
each setting is correct and restart pump running.
NOTE: Rate, dose, and feed interval settings can be locked into place
so that changes cannot be made. See the User Preference Settings
section (page 25) for instructions to change this feature.
22
Infinity
Enteral Feeding Pump Operator's Manual
®