Moog infinity Manuel De L'opérateur page 20

Masquer les pouces Voir aussi pour infinity:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

2. DIRECTIONS FOR USE
RUN/PAUSE
To restart pump when in paused mode, press the
key. The
pump will restart the feeding at the point where it was paused.
NOTE: ANY CHANGES TO PUMP SETTINGS DURING A FEEDING CYCLE WILL CAUSE
PUMP TO START A NEW FEEDING; IT WILL NOT START WHERE IT LEFT OFF.
Feeding Completion:
When the dose has been completely delivered, pump will stop running, and display
will read DOSE DONE. Pump will be in a pause mode until it is turned off or feeding
ON/OFF
is restarted. To clear DOSE DONE, press and hold the
key for 1.5 seconds to
turn pump off, or press the
RUN/PAUSE
key to put pump in pause mode. Adjustments
RUN/PAUSE
to settings can be made at this time, or press the
key a second time
to start another feeding using same rate, dose, and feed interval settings.
If you would prefer the alarm to sound when the
dose is complete, you can change the pump dose
complete alarm setting to BEEP WHEN DONE. See
the User Preference Settings section (page 25)
for instructions to change this setting. When set to
BEEP WHEN DONE, the pump will stop at the end
of the feeding, beep intermittently, and display
will read DOSE DONE (Figure 2-18). To silence alarm
ON/
when the dose is complete, press and hold the
OFF
key for 1.5 seconds to turn pump off, or press
RUN/PAUSE
the
key to put pump in pause mode.
Figure 2-18 DOSE DONE
NOTE: To program an infinite dose, press
+
and hold the
key until display reads INF.
When an infinite dose has been programmed, pump will not alarm DOSE
DONE. Pump will run continuously as long as feeding solution remains in
the bag. When empty, pump will alarm NO FOOD or NO FLOW IN.
18
Infinity
Enteral Feeding Pump Operator's Manual
®

Publicité

Table des Matières
loading

Table des Matières