2. DIRECTIONS FOR USE
Step 5:
Loop silicone tubing around pump wheel
stretching lightly. Place cassette into pump
(Figure 2-8). Close pump door (Figure 2-9).
NOTE: See pages 40-41 for
pump use with a backpack.
Step 6:
ON/OFF
Press and hold
key for 1.5 seconds to
Figure 2-8 Seat Cassette
turn pump on. While pump runs through a self
test, display will light and an audible alarm will
sound as pump displays the nine digit serial
number three digits at a time for one second each.
The display will then show the letter 'R' followed
by a number, which is the software revision.
Next, all segments of display will be
shown for 2 seconds. Verify all display
segments and symbols are active.
Figure 2-9 Close Pump Door
The self test is complete and pump will then display
last programmed rate and will be in pause mode.
If any air is still in the tubing, continue to
Step 7 to use pump prime feature.
Step 7:
PRIME
Press and hold
key. Alarm will sound
once and pump will begin pumping at an
approximate rate of 700 ml/hr. Display will read
Figure 2-10 Priming Indication
"TO STOP LET GO" (Figure 2-10). Once all air is
removed from tubing, release key. Pump will stop,
display will revert to last programmed rate, and pump will be in pause mode.
NOTE: Occlusion and air in line alarms are disabled while priming.
NOTE: Pump also has the ability to auto prime. To enable the auto
prime function, see User Preference Settings on page 26.
For pump operation instructions:
4 For a Single Feeding Example: go to page 16.
4 For an Interval Feeding Example: go to page 19.