2. DIRECTIONS FOR USE
Step 4:
+
-
Press the
or
key to change the rate to 50 ml/hr.
Hold down either key to change rapidly (Figure 2-19).
Step 5:
RATE/DOSE
Press the
or
-
key until a dose of 100 ml is displayed (Figure 2-20).
Step 6:
FEED INT
Press the
interval setting. Press the
6.00 hr is displayed (Figure 2-21).
NOTE: The amount of time to complete a rate and
dose combination can be figured by dividing the
dose by the rate (Dose/Rate = Time). The number of
hours to be programmed as the feed interval is the
amount of time to deliver the dose plus the amount
of time pump should pause. Example: 100 ml divided
by 50 ml/hr equals 2 hours for feeding delivery, plus a
pause of 4 hours equals a feeding interval of 6 hours
(100 ml ÷ 50ml/hr = 2 hours running + 4 hours pause
= 6 hours from feeding start time to start time).
Pump will calculate number of hours it will take to
deliver the rate and dose combination. When the
FEED INT
key is pressed the first available value
will be compatible with that combination. In this
example the first available value will be 2.00 hr.
NOTE: Feed interval is displayed in hours and
minutes, i.e. 5.10 = 5 hours and 10 minutes.
NOTE: Feed interval cannot be programmed
if rate and dose combination requires
more than 24 hours to deliver.
Step 7:
Connect end of disposable set to patient's enteral feeding tube. (Figure 2-22).
Infinity
Enteral Feeding Pump Operator's Manual
®
key to display dose. Press the
key to display the feed
+
or
-
key until
+
20
Figure 2-19 Program
Rate to 50 ml/hr
Figure 2-20 Program
Dose to 100 ml
Figure 2-21 Program
Feed Interval 6.00 hr
Figure 2-22 Connect disposable
set to feeding tube