3. ALARMS, MESSAGES, AND INDICATIONS
NO FLOW IN (Figure 3-4)
Pump will sound a dual tone beep repeatedly
and display will read NO FLOW IN.
Why: This alarm has occurred because
pump has detected blockage in the
set between pump and bag.
To correct the problem: Press the
key to silence alarm and place pump in
pause mode, then check the following:
1. Check tubing for kinks or formula
clumps. Correct blockage.
2. Check cassette receptacle for cleanliness,
especially around the pressure sensors.
If cleaning is necessary, refer to
CHAPTER 5 - CLEANING (page 38).
NO FLOW OUT (Figure 3-5)
Pump will sound a dual tone beep repeatedly
and display will read NO FLOW OUT.
Why: This alarm has occurred because
pump has detected a blockage in the
set between pump and patient.
To correct the problem: Press the
key to silence alarm and place pump in
pause mode, then check the following:
1. Check tubing for kinks or blockages.
Remove kink or blockage.
2. Check cassette receptacle for cleanliness,
especially around pressure sensors. If
cleaning is necessary, refer to CHAPTER
5 - CLEANING (page 38).
3. Back pressure from patient may also cause this alarm. Discuss with physician.
Infinity
Enteral Feeding Pump Operator's Manual
®
RUN/PAUSE
RUN/PAUSE
30

Figure 3-4 NO FLOW IN Alarm

Figure 3-5 NO FLOW
OUT Alarm