4. TROUBLESHOOTING GUIDE
ER01 - ER99 Alarms
Was pump door open
while pump was turned
on or while running?
NO FLOW IN or NO FLOW OUT Alarms
4 NO FLOW IN - Occlusion between delivery set and pump.
4 NO FLOW OUT - Occlusion between pump and patient.
Is delivery set tubing
pinched, kinked,
or clogged?
NO
Is tab on inside of
door broken?
NO
Is pressure sensor
region of cassette
receptacle clean?
YES
Are there visible signs
of damage to pressure
sensors area?
Infinity
Enteral Feeding Pump Operator's Manual
®
YES/UNSURE
NO
Check delivery set for obstructions or kinks in
YES
tubing. Correct blockage and resume feeding.
Contact Moog Customer Service at
YES
800.970.2337 to order a new door.
Clean pressure sensor area with a cotton swab, soft cloth,
NO
or dampened sponge; or wash entire pump under running
water. Do not use abrasive materials or harsh chemicals.
Contact Moog Customer Service at
YES
800.970.2337 to return pump for service
Disconnect from patient and fill a new delivery
set with food. Prime set and start pump.
NO
continues, contact Moog Customer Service at
800.970.2337 to return pump for assistance.
36
Turn pump off. Unload and reload delivery
set cassette. Make sure door is closed
securely, then turn pump back on.
If alarm continues, contact Moog Customer
Service at 800.970.2337 for assistance.
If alarm