Alarms

ER01 - ER99 (Figure 3-1)
Pump will sound a continuous beep and display
will read ER followed by a two digit number or ERR
followed by a letter to indicate a self-test has failed.
To clear alarm: Turn the pump
off and then back on.
WARNING: If an error occurs
(ER01 - ER99 will appear in display), all settings and volume counters
should be checked and verified before starting a feeding.
LOAD SET (Figure 3-2)
Pump will sound a dual tone beep repeatedly
and display will read LOAD SET.
Why: This alarm will sound when pump attempted
to run with the set improperly loaded or missing.
To correct the problem: Press the
key to silence alarm and place pump in
pause mode, then check the following:
1. An Infinity disposable set is loaded into
pump properly and door is closed.
2. Check cassette receptacle for cleanliness,
especially around pressure sensors. If
cleaning is necessary, refer to CHAPTER
5 - CLEANING (page 38).
LOW BATT (Figure 3-3)
When 1 hour or less of battery life is available,
display will flash LOW BATT every 3 seconds and
pump will beep every 2 seconds to indicate battery
is low and pump will soon stop running.
To correct the problem: Plug in AC
Adapter/Charger to recharge battery.
3. ALARMS, MESSAGES, AND INDICATIONS
RUN/PAUSE
29
Figure 3-1 ER01 - ER99 Alar ms

Figure 3-2 LOAD SET Alarm

Figure 3-3 LOW BATT Alarm

© Copyright 2022 Moog, Inc.