ALARMS
ACCESSORIES
WARRANTY
OPERATING
ENVIRONMENT
STORAGE
ENVIRONMENT
SERVICE LIFE
CLASSIFICATION
INFORMATION
CHEK
ER01 - ER99
LOAD SET
LOW BATT
NO FLOW IN
NO FLOW OUT
NO FOOD
PUSH RUN TO FEED
SHUT DOOR
AC Adapter/Charger
Multi-position Pole Clamp (rotates 360°)
Multi-purpose Infinity Custom Carry Packs
2 years
Temperature:
Humidity:
Pressure:
Temperature:
Humidity:
The pump is designed to provide a minimum of 5 years of service life
Infinity Enteral Feeding Pump with internal battery
Type BF Applied Part
Infinity AC Adapter / Charger – Class II
Programming error - Incompatible RATE,
DOSE, and FEED INTERVAL settings
Self-test error
Set not loaded properly
Battery Low
Upstream occlusion
-5 psi (-34 kPa)
Tolerance = ± 3psi (21 kPa)
psi (pounds per square inch)
kPa (kilopascals)
Downstream occlusion,
12 psi (83 kPa)
Tolerance = ± 3psi (21 kPa)
psi (pounds per square inch)
kPa (kilopascals)
Downstream occlusion will alarm at
5 minutes at 125 ml/hr
2 hours at 1 ml/hr
12 hours at 0.1 ml/hr
After occlusion is cleared, up to 1
ml of fluid will be released
Air in tubing - bag or set may be empty
Pump unused for 2 minutes
Door opened while pump was running
41 °F to 104 °F (5 °C to 40 °C)
10% to 95%, non- condensing
700 -1,060 hPa
These conditions apply before and after the pump and
disposable have been removed from their packaging.
The pump requires 30 minutes to warm up/
cool down to its normal operating temperature
from low/high storage temperatures.
-4 °F to 122 °F (-20 °C to +50 °C)
10% to 95% noncondensing
These conditions apply before and after the pump and
disposable have been removed from their packaging.
43
7. SPECIFICATIONS
© Copyright 2022 Moog, Inc.