5. CLEANING
5. CLEANING
WARNING: To avoid electrical shock, never clean pump or Infinity AC Adapter/
Charger with charger plugged into an outlet or pump turned on.
WARNING: Make sure the Infinity AC Adapter/Charger is
completely dry before plugging into an electrical outlet.
To clean the Infinity enteral feeding pump:
Pump may be cleaned with warm soapy water (standard dish soap) and a
nonabrasive sponge or soft cloth. Use a cotton swab to clean pathways of cassette
receptacle and to remove teal silicon residue from pump wheel roller pins.
OR - Apply one of the following solutions for approximately 10
minutes, then wipe pump clean with a damp cloth or sponge:
4 5% bleach and water solution
4 Multipurpose household disinfectant cleaner
Rinse pump by holding under a stream of
warm water. Then, dry with a clean cloth.
NOTE: Clean pump as needed. If the pump is used on
more than one patient, clean between each patient.
NOTE: Avoid harsh cleaners/disinfectants. They
may cause damage to pump surfaces and may
affect pump functionality. Moog has tested
the following disinfectants for acceptability for
cleaning/disinfecting the Infinity pump. These
cleaning products are categorized below:
Acceptable
Metrex CaviCide
5% Bleach and Water Solution
Warm Soapy Water (Standard Dish Soap)
To clean the Carrying Packs:
The Infinity packs are machine-washable. Use cold water
and gentle cycle, whenever possible. Hang to dry.
To clean the AC Adapter/Charger:
Infinity AC Adapter/Charger normally does not require cleaning. When
desired, a dry or slightly damp cloth may be used to clean the outside surface
of AC Adapter/Charger while it is disconnected from the wall outlet.
Infinity
Enteral Feeding Pump Operator's Manual
®
Unacceptable
Ball Ruthless
Wex-Cide 128
Alcohol
Vesphene Ilse
38
Figure 5-1 Rinsing Pump
Under Stream of Water