• Do not use the patient helper as a push or pull device.
• Do not use the oxygen bottle holder as a push or pull device.
• Do not use the IV pole as a push or pull device.
• Always make sure that the IV pole is at a low height during transport.
• Always make sure that all persons and equipment are away from the area below and around the Fowler before you
activate the CPR release. The CPR release is for emergency use only.
• Always expand or retract the litter width before you place a patient on the product.
• Always expand or retract the litter width when the product is in the full flat horizontal position.
• Always expand the litter deck to the desired width before you expand the headboard.
• Always retract the headboard to the desired width before you retract the litter deck.
• Do not expand or retract the headboard while a patient is on the product.
• Always make sure that the product is clear of obstacles before you use motion functions.
• Always raise the siderails when the litter is in its full down position. This prevents the scale system from weighing a
patient inaccurately.
• Do not load the IV pole above the safe working load of 40 lb (18 kg).
• Do not load an individual IV pole hook above the safe working load of 20 lb (9 kg).
• Always secure the lifting pole in the mounting bracket before you adjust the patient helper.
• Always make sure that the patient helper mounting bracket is secure before use.
• Do not load the oxygen bottle holder above the safe working load of 45 lb (20.4 kg).
• Always unplug the power cord from the wall outlet when large spills occur near the circuit boards, cables, and motors.
Remove the occupant from the product, clean up the fluid, and inspect the product. Fluids can cause unpredictable
operation and decreased functionality of any electrical product. Do not return the product to service until dry and tested
for safe operation.
• Always wipe down with clean water (or 70% isopropyl alcohol, if using V V i i r r e e x x ® TB) and dry each product after
disinfecting. Some disinfectants are corrosive in nature and may cause damage to the product. If you do not rinse and
dry the product, you may leave a corrosive residue on the surface of the product. This corrosive residue could cause
premature degradation of critical components. Failure to follow these disinfecting instructions may void your warranty.
5900-009-001 Rev F.1
5