D D i i s s i i n n f f e e c c t t i i n n g g
C C A A U U T T I I O O N N
• Do not clean, disinfect, service, or perform maintenance while the product is in use.
• Always unplug the power cord from the wall outlet when large spills occur near the circuit boards, cables, and motors.
Remove the occupant from the product, clean up the fluid, and inspect the product. Fluids can cause unpredictable
operation and decreased functionality of any electrical product. Do not return the product to service until dry and tested
for safe operation.
• Always wipe down with clean water (or 70% isopropyl alcohol, if using V V i i r r e e x x ® TB) and dry each product after
disinfecting. Some disinfectants are corrosive in nature and may cause damage to the product. If you do not rinse and
dry the product, you may leave a corrosive residue on the surface of the product. This corrosive residue could cause
premature degradation of critical components. Failure to follow these disinfecting instructions may void your warranty.
Recommended disinfectants for this product's surfaces include the following:
Quaternary (active ingredient - ammonium chloride) that contain ≤ 3% glycol ether
Phenolic (active ingredient - o-phenylphenol)
Chlorinated bleach solution (5.25% - less than 1 part bleach to 100 parts water)
Alcohol (active ingredient - 70% isopropyl alcohol)
Recommended disinfection method:
1. Follow the disinfectant solution manufacturer's dilution recommendations.
2. Apply the recommended disinfectant solution by spray or pre-soaked wipes.
3. Hand wash all exposed surfaces of the product with the recommended disinfectant.
4. Dry the product before you return it to service.
N N o o t t e e
• Avoid oversaturation. Do not allow the product to remain wet.
• Follow the manufacturer's dilution recommendations for appropriate contact time and rinsing requirements. Follow the
chemical manufacturer's guidelines for proper disinfecting.
EN
44
5900-009-001 Rev F.1