• M M V V 3 3 is equipped with inputs that accept either a DB-37 or 1/4" jack nurse call cable.
To setup nurse call communication:
1. Plug the interface cable into the 37-pin connector or 1/4" jack in the litter frame at the head end of the product (A)
(Figure 4) and into the applicable connection (patient station, head wall, or docker station).
N N o o t t e e
• Only connect the 37-pin connector or 1/4" jack to the head wall output configuration or product Communications
Tester (sold separately).
• Make sure that you insert the 1/4" jack dummy plug (C) into the 1/4" jack when not in use.
2. Use a #2 Phillips screwdriver and two screws (#10-32) to secure the strain relief (B) that holds the interface cable to the
litter frame for strain relief (Figure 4).
3. Press the N N u u r r s s e e c c a a l l l l button (H, Operator control panel (outside siderail) (page 25)) to verify the connection between the
product's nurse call signal and the hospital's nurse call system.
To activate nurse call, see Activating nurse call communication (page 25).
5900-009-001 Rev F.1
B
A
F F i i g g u u r r e e 4 4 – – 3 3 7 7 - - p p i i n n c c o o n n n n e e c c t t o o r r a a n n d d 1 1 / / 4 4 " " j j a a c c k k
C D
13