W W e e i i g g h h i i n n g g a a p p a a t t i i e e n n t t
W W A A R R N N I I N N G G
• Do not use the scale system reading as a reference for medical treatment.
• The scale system assists only in the monitoring of the patient's weight variation.
• Do not use the scale system on an unlevel floor greater than 2.86°.
C C A A U U T T I I O O N N - - Always raise the siderails when the litter is in its full down position. This prevents the scale system from
weighing a patient inaccurately.
Before you place a patient on the product, make sure that you zero/tare the scale (see Zeroing/taring the scale (page 32)).
N N o o t t e e - - Always zero/tare the scale after you add a support surface, mattress, or linens to the product.
To weigh a patient:
1. Press the scale button (F) on the footboard control panel (see Footboard control panel - menu controls (page 30)).
2. On the scale display, use the up arrow button (D) and down arrow button (E) to highlight the W W e e i i g g h h P P a a t t i i e e n n t t option (see
Footboard control panel - menu controls (page 30)).
3. When the W W e e i i g g h h P P a a t t i i e e n n t t option says R R e e a a d d y y , press the enter button (C) (see Footboard control panel - menu controls
(page 30)).
N N o o t t e e - - Do not touch the product when you weigh the patient.
The W W e e i i g g h h P P a a t t i i e e n n t t option will display the patient's weight for a moment before it is displayed on the next line, L L a a s s t t w w e e i i g g h h t t .
N N o o t t e e - - If a previous weight was displayed on the L L a a s s t t w w e e i i g g h h t t line, it will now appear in W W e e i i g g h h t t H H i i s s t t o o r r y y .
A A d d d d i i n n g g o o r r r r e e m m o o v v i i n n g g e e q q u u i i p p m m e e n n t t
To change equipment:
1. Press the scale button (F) on the footboard control panel (see Footboard control panel - menu controls (page 30)).
2. On the scale display, use the up arrow button (D) and down arrow button (E) to highlight the C C h h a a n n g g e e E E q q u u i i p p m m e e n n t t option
(see Footboard control panel - menu controls (page 30)).
3. Press the enter button (C) and follow the prompts to add or remove equipment (see Footboard control panel - menu
controls (page 30)).
V V i i e e w w i i n n g g t t h h e e w w e e i i g g h h t t h h i i s s t t o o r r y y
W W A A R R N N I I N N G G - - Do not use the product clock to replace patient monitoring protocol, the product clock is for reference only.
To view the weight history log:
1. Press the scale button (F) on the footboard control panel (see Footboard control panel - menu controls (page 30)).
2. On the scale display, use the up arrow button (D) and down arrow button (E) to highlight the W W e e i i g g h h t t H H i i s s t t o o r r y y option (see
Footboard control panel - menu controls (page 30)).
3. Press the enter button (C) to view the weight history (see Footboard control panel - menu controls (page 30)).
The weight history stores up to 10 scale readings.