P P r r e e v v e e n n t t i i v v e e m m a a i i n n t t e e n n a a n n c c e e
Remove the product from service before you perform the preventive maintenance inspection. Check all items listed during
annual preventive maintenance for all Stryker Medical products. You may need to perform preventive maintenance checks
more often based on your level of product usage. Service only by qualified personnel.
N N o o t t e e - - Clean and disinfect the exterior of the support surface before inspection, if applicable.
Inspect the following items:
All welds
All fasteners are secure
Casters lock with brake pedal applied
Casters are secure and swivel
Casters are free of wax and debris
Footboard light LED and bumper LEDs flash when brakes are released
Steer function works
Siderails move, latch, and stow
Deck expands, retracts, and locks in all positions
CPR release operates
IV pole is intact and operates (option)
Foley bag hooks intact
No cracks or splits in headboard, footboard or siderail panels
No rips or cracks in mattress cover
All functions on head end siderails operate (includes LEDs)
All functions on footboard operate (includes LEDs)
No cracks or damage to control overlays
Calibrate scale system
Night light operates
Power cords and plug not frayed or damaged
No damage to nurse call connections
All ground strap cables are secure to the frame
Ground impedance not more than 200 mΩ (milliohms)
Current leakage not more than 300 µA (microamps)
Check lift actuator clevis pins for grease, apply if needed
Ground chains intact
Trendelenburg/reverse Trendelenburg angle accuracy is minimum +12° to -12°
Fowler angle accuracy is 0° - 60° ± 2°
Fowler holds position at 30° with weight
Siderail switches operate ( i i B B e e d d Awareness)
Footboard light LED and bumper LEDs operate ( i i B B e e d d Awareness)
Inspect footboard control labels for signs of degradation
Inspect hi-lo actuators for oil leaks
Inspect footboard connector housing for cracks or damage
All motions function
5900-009-001 Rev F.1