Stryker MV3 Bariatric Bed Mode D'emploi page 21

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

N N o o t t e e - - To move the product in any direction, including laterally, release the steer lock pedal.
F F i i g g u u r r e e 8 8 – – A A p p p p l l y y i i n n g g s s t t e e e e r r l l o o c c k k
A A c c t t i i v v a a t t i i n n g g t t h h e e C C P P R R r r e e l l e e a a s s e e
C C A A U U T T I I O O N N - - Always make sure that all persons and equipment are away from the area below and around the Fowler before
you activate the CPR release. The CPR release is for emergency use only.
When you raise the Fowler and need quick access to the patient, you can position the product Fowler angle to 0° by
activating the CPR release.
You can find the two CPR release levers at the head end section on both the left and right sides of the Fowler (A) (Figure
10).
To activate the CPR release:
1. Pull the lever (A) on either side of the Fowler (Figure 10).
N N o o t t e e - - Release the CPR lever at any time to stop product Fowler, knee Gatch, and foot section motion.
2. Guide the Fowler to the flat position.
N N o o t t e e
• The knee Gatch and foot section will lower when you pull the CPR lever.
• You can also activate the CPR release if you press the C C P P R R button on the footboard control panel ( Footboard control
panel - bed controls (page 27)).
5900-009-001 Rev F.1
F F i i g g u u r r e e 9 9 – – R R e e l l e e a a s s i i n n g g s s t t e e e e r r l l o o c c k k / / n n e e u u t t r r a a l l p p o o s s i i t t i i o o n n
17
EN

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

5900