S S e e t t u u p p
W W A A R R N N I I N N G G - - Always plug the product into a grounded, hospital grade wall outlet. You can only achieve grounding reliability
when you use a hospital grade wall outlet. This product is equipped with a hospital-grade plug for protection against electric
shock hazard.
C C A A U U T T I I O O N N - - Do not clean, disinfect, service, or perform maintenance while the product is in use.
N N o o t t e e - - Allow the product to reach room temperature before you conduct any setup or test functional operations.
Before you place the product into service, make sure these components are working properly:
1. Plug the product into a grounded, hospital grade wall outlet and make sure that the power LED at the foot end of the
product comes on ( Plugging or unplugging the product (page 14)).
2. Make sure that the deck extensions expand, retract, and lock in the 36", 42", and 48" positions ( Expanding or retracting
the litter (page 18)).
3. Make sure that the siderails raise, lower, lock in the up position, lock in the intermediate position when lowered and store
( Raising the siderails (page 21)).
4. Make sure that all four casters lock when you apply the brake( Applying or releasing the brakes (page 16)).
N N o o t t e e - - Make sure that the B B r r a a k k e e LED located on the footboard control panel ( Footboard control panel - home display
(page 31)) illuminates when the brakes are applied and blinks when the brakes are released.
5. Raise the Fowler (head of bed) up to approximately 60°. Pull the CPR release handle and make sure that the back will
drop with minimal effort ( Activating the CPR release (page 17)).
6. Perform each function on the footboard control panel to make sure that each function works.
7. Perform each function on each control panel on the head end siderails to make sure that each function works.
8. Activate the motion stop system. Press B B e e d d h h e e i i g g h h t t d d o o w w n n (J, Footboard control panel - bed controls (page 27)) to lower
the litter. As the litter lowers, push up on the motion interrupt pan under the litter and make sure that downward motion
stops. Release the pan and allow downward motion to continue
N N o o t t e e - - The product's upward motion or other functions are not disrupted by the motion stop system.
9. Set clock to local date and time ( Setting the clock (page 36)).
10. Check or set zone of operation to area gravity zone (for NAWI scale only). Service only by qualified personnel.
A A t t t t a a c c h h i i n n g g t t h h e e b b a a r r i i a a t t r r i i c c n n o o n n - - p p o o w w e e r r e e d d s s u u p p p p o o r r t t s s u u r r f f a a c c e e ( ( o o p p t t i i o o n n ) )
To attach the bariatric non-powered support surface option onto M M V V 3 3 , see the Bariatric Non-Powered Support Surface
Operations Manual (2885-009-001).
N N o o t t e e - - Always zero/tare the scale after adding a support surface or mattress to the bed frame, see Zeroing/taring the scale
(page 32).
S S e e t t t t i i n n g g u u p p n n u u r r s s e e c c a a l l l l c c o o m m m m u u n n i i c c a a t t i i o o n n
W W A A R R N N I I N N G G - - Always use a Stryker supplied interface cable. Use of any other cable may cause the product to not function
as intended, which may result in patient or user injury.
N N o o t t e e
• The nurse call button on this product has not been evaluated for the requirements of Clause 17 (Normal Operation) of
UL 1069. The user is responsible to determine the operability of the nurse call button with all systems connected to the
product.
• The nurse call interface connects to nurse call systems designed to comply with appropriate standards (i.e. IEC 60601-
1, IEC 60950, UL 1069) and rated for a maximum of 42.4Vpk, 60Vdc.
EN
12
5900-009-001 Rev F.1