C C l l e e a a n n i i n n g g
C C A A U U T T I I O O N N
• Do not clean, disinfect, service, or perform maintenance while the product is in use.
• Always unplug the power cord from the wall outlet when large spills occur near the circuit boards, cables, and motors.
Remove the occupant from the product, clean up the fluid, and inspect the product. Fluids can cause unpredictable
operation and decreased functionality of any electrical product. Do not return the product to service until dry and tested
for safe operation.
Recommended cleaning method:
1. Hand wash all exposed surfaces of the product with a mild detergent by spray or pre-soaked wipes.
2. Follow the cleaning solution manufacturer's instructions for appropriate contact time and rinse requirements.
3. Dry the product before you return it to service.
N N o o t t e e - - Avoid oversaturation. Do not allow the product to remain wet.
5900-009-001 Rev F.1