Stryker MV3 Bariatric Bed Mode D'emploi page 29

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

A A c c t t i i v v a a t t i i n n g g n n u u r r s s e e c c a a l l l l c c o o m m m m u u n n i i c c a a t t i i o o n n
Nurse call allows the patient or healthcare professional to send a signal to the nurse station for assistance.
To activate nurse call, press the N N u u r r s s e e c c a a l l l l button (H, see Operator control panel (outside siderail) (page 25)) (A, Patient
control panel (inside siderail) (page 26)).
N N o o t t e e - - Nurse call requires a connection between M M V V 3 3 and an applicable input (patient station, head wall, or docker
station). See Setting up nurse call communication (page 12).
C C o o n n n n e e c c t t i i n n g g p p e e r r i i p p h h e e r r a a l l e e q q u u i i p p m m e e n n t t t t o o t t h h e e h h o o s s p p i i t t a a l l g g r r a a d d e e G G F F C C I I a a u u x x i i l l i i a a r r y y o o u u t t l l e e t t
W W A A R R N N I I N N G G
• Only use hospital grade electric equipment consuming 6A (120 VAC)/3A (230 VAC) or less with the hospital grade GFCI
auxiliary outlet. The use of standard electric equipment may bring the current leakage to a level unacceptable for
hospital equipment.
• Do not use the 120/230 VAC hospital grade GFCI auxiliary outlet for life sustaining equipment.
The 120/230 VAC hospital grade GFCI auxiliary outlet is a built-in outlet for peripheral equipment. You can find the outlet at
the foot end of the product ( Product illustration (page 9)).
N N o o t t e e - - The hospital grade GFCI auxiliary outlet is protected by resettable circuit breakers located at the head end of the
product (D, Figure 4).
O O p p e e r r a a t t o o r r c c o o n n t t r r o o l l p p a a n n e e l l ( ( o o u u t t s s i i d d e e s s i i d d e e r r a a i i l l ) )
W W A A R R N N I I N N G G
• Always lower the product to its lowest height when the patient is unattended to reduce the risk of injury due to patient
falls.
• Always make sure that cables, wires, and tubing from other equipment are routed so that they are not pinched by parts
of the product.
5900-009-001 Rev F.1
A
F F i i g g u u r r e e 1 1 9 9 – – F F o o l l e e y y b b a a g g h h o o o o k k
25
EN