O O p p e e r r a a t t i i o o n n
P P l l u u g g g g i i n n g g o o r r u u n n p p l l u u g g g g i i n n g g t t h h e e p p r r o o d d u u c c t t
W W A A R R N N I I N N G G
• Always connect this product to a supply mains with protective earth to avoid the risk of electric shock.
• Always make sure the product is connected to an appropriate power source if the loss of power would result in
unacceptable risk.
• Always store the power cord to avoid the risk of entanglement, damage to the power cord, or potential shock hazards. If
the power cord is damaged, remove the product from service and contact maintenance personnel.
• Always allow enough clearance between the head end of the product and the adjacent wall, so you can unplug the
power cord from the wall outlet during an emergency.
• Always store the power cord before you transport the product.
To plug in the product, plug the power cord into a hospital grade protective earthed outlet.
To unplug the product, grasp the mold near the outlet and pull in a direction parallel to the floor (not at an angle).
N N o o t t e e - - The P P o o w w e e r r LED (K) on the footboard control panel ( Footboard control panel - menu controls (page 30)) illuminates
when the product is plugged in.
C C h h a a r r g g i i n n g g t t h h e e b b a a t t t t e e r r y y
W W A A R R N N I I N N G G
• Always disconnect the power cord from the wall outlet if you detect overheating of the battery, cables, or cords. Do not
use the product until it has been inspected, serviced, and confirmed to work as intended by maintenance personnel.
• Always replace the battery after it surpasses its expected service life.
• Do not open the battery.
• Do not expose the battery to excessive heat.
• Do not spill liquid onto the battery or submerge the battery in liquid.
C C A A U U T T I I O O N N
• Always plug the product into a wall outlet (regulated AC-power source) when not in use to maintain a sufficient battery
charge and to maximize product performance while operating on battery power.
• Always replace batteries that have corrosion at the terminals, display cracking, have expanded or bulging sides, or no
longer can maintain a full charge.
• Always use authorized batteries when you replace the batteries. Use of unauthorized batteries may lead to
unpredictable system performance.
M M V V 3 3 is equipped with a battery backup system that charges when the product is plugged into a wall outlet. The battery
backup system allows the operator to use the product when the product is unplugged, during a power failure, or while
transporting a patient. Battery backup functionality activates when you unplug the product.
Always check battery backup function according to the preventive maintenance checklist (see Preventive maintenance
(page 45)). Always replace the battery if it does not perform as intended during preventive maintenance.
To charge the battery, connect the product to a wall outlet. The battery has a full charge within 10 to 12 hours.
N N o o t t e e
• The P P o o w w e e r r LED (K) and B B a a t t t t e e r r y y LED (J) on the footboard control panel ( Footboard control panel - menu controls (page
30)) illuminate when the product is plugged in and the battery is charging.
• The B B a a t t t t e e r r y y LED flashes when the product is on battery power.
• While on battery power, the B B a a t t t t e e r r y y LED will flash slowly when full and flash fast when low.