Setting The Zero Point (Offset) (Off.p); Automatic Offset Setting; Edited Offset Setting - JUMO DELOS HP Notice De Mise En Service

Masquer les pouces Voir aussi pour DELOS HP:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Display and setting possibilities of the device
Measuring range
0 to 160 bar
0 to 600 bar
7.5

Setting the zero point (offset) (Off.P)

7.5.1

Automatic offset setting

This setting applies the current measured value as the new zero point.
1. Unlock the device.
 chapter 7.2 "Unlocking the device (enter the code).", page 23
2. "Turn" until "Off.P" appears in the bottom line.
3. "Press" twice in quick succession. This applies the current measured value as the zero point.
7.5.2

Edited offset setting

This setting lets you increase or reduce the measured pressure by a configurable value.
1. Unlock the device.
 chapter 7.2 "Unlocking the device (enter the code).", page 23
2. "Turn" until "Off.P" appears in the bottom line.
3. "Press"
NOTE!
"-" means: the offset is negative - the measured pressure is reduced.
The value is entered "digit-by-digit".
Unit
bar
MPa
psi
%
bar
MPa
psi
%
Display
Start
0.0
0.00
0
0.0
0.0
0.00
0
0.0
7 Startup
End
160.0
16.00
2321
100.0
600.0
60.00
8702
100.0
25