Error message
Cause
Gone above measured value
The measured value is too large, is
outside the measurement range, or a
probe break has occurred.
Gone below measured value
The measured value is too small, is
outside the measurement range, or a
probe short-circuit has occurred.
Display of measurement value 2 (-
10.5°C) alternating with background
information that measurement value 1 is
faulty..
.Display of measurement value 1 (-
18.5°C) alternating with background
information that measurement value 2 is
faulty..
Measured value connot be displayed
The measured value exceeds 999 or is
below -999 and is thus outside the
3-digit display range.
8 Alarm and error messages
Remedy
-
Check probe and connecting cable for damage or
short-circuit
-
Check whether the correct probe has been set or
connected
v
Chapter 4 "Commissioning the instrument"
H
These messages are only output
in the process value display.
-
Change over to the faulty measurement value and
find the cause of the error (see above)
-
Analog input 2 must be activated and correctly
configured.
v
Section 5.1 „Analog inputs"
-
If necessary, use the setup program to re-transfer
the configuration to the instrument!