Determining Density Of Liquids - KERN ABT Serie Mode D'emploi

Masquer les pouces Voir aussi pour ABT Serie:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Repeatedly press the [UNIT] key until the balance is in density
determination mode for solids ",d". Note that "g" also appears dur-
ing weight measurement in air.
Press the [TARE] key. Place the item to be measured on the
weighing pan.
When standstill control is complete, press the [CAL] key
Place the item to be measured on the immersed hanging weighing
pan. The display is showing the density of the measured item.
"dSP oL" may be displayed when nothing is on the weighing pan,
which is normal.

14.3.2 Determining density of liquids

Condition: Function "d" (density of liquids) is activated. See Chapter 11.3.
Press the [CAL] key repeatedly until "SettinG" appears.
Press the [TARE] key.
Repeatedly actuate the [CAL] key until "Sv SEt" appears
Press the [TARE] key. The currently set density for the body to be
immersed appears. In the upper part of the display panel, the
(Example)
ical input status. The leftmost digit blinks.
Enter density for your body to be immersed. When the [UNIT] key
is pressed, the numerical of the blinking digit increases by 1 at a
(Example)
time. You can determine the value of the flashing digit, or shift the
flashing digit by one position to the right, by pressing the [PRINT]
key. Confirm your setting by pressing the [TARE] key.
Repeatedly press the [ON/OFF] key until the balance is in weigh-
ing mode.
ABT-NM-BA-e-1622
symbol and the # symbol appear in order to indicate numer-
57

Publicité

Table des Matières
loading