Iso/Glp Log; Setting The Log Printout Of Your Adjustment Data - KERN ABT Serie Mode D'emploi

Masquer les pouces Voir aussi pour ABT Serie:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

10 ISO/GLP log

Quality assurance systems require printouts of weighing results as well as of correct
adjustment of the balance stating date and time and balance identification. The easi-
est way is to have a printer connected.

10.1 Setting the log printout of your adjustment data

This function enables automatic log issue after each adjustment. These logs may be
issued by using an optional printer. Fully automatic and periodical adjustments are
carried out and logs created in combination with the Clock-CAL function (see ch.
8.2).
First make adjustments as specified in chap. 15.4 under „KERN-YKB-01N".
Then set the log printout as follows:
Press the [CAL] key repeatedly until "SettinG" appears.
Press the [TARE] key. The display shows "CAL dEF".
Repeatedly actuate the [CAL] key until "GLP:**" appears (The ∗∗
symbolise the current setting).
To change settings, actuate the [TARE] key when "GLP:**" ap-
pears.
You can use the [CAL] key to select between the following set-
tings:
"GLP-on"
"GLP-oF"
The current setting is indicated by the standstill display ( ).
Confirm your selection by pressing the [TARE] key
Actuate the [ON/OFF] key repeatedly. Return to menu/weighing
mode.
ABT-NM-BA-e-1622
Function activated
Function deactivated
31

Publicité

Table des Matières
loading

Table des Matières