Add Up Mode - KERN ABT Serie Mode D'emploi

Masquer les pouces Voir aussi pour ABT Serie:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

14.5 Add up mode

An optional number of single weighing processes are automatically added up to a
total sum, e.g. all single weighing processes of a batch or if you have to carry out a
great number of measurements of small samples.
When the standstill control ( ) is complete the weighing value is automatically is-
sued to the optional printer. The displayed value is added into the total adding
memory. Afterwards automatic taring will take place. This is repeated newly every
subsequent time a new sample is placed on the weighing pan. When the last single
weighing process is finished, press the [ON/OFF] key to receive the total sum ("TO-
TAL=").
Important:
 This mode can be applied to any unit.
 The initial load of the balance must be kept within the ± 5-fold of the ze-
ro range.
 Weight of sample ³ 10 x zero range (zero range = 0.25 e)
 If during the use of multiple range balances the smallest readability digit
was turned off by pressing the [1d/10d] key, evaluation is based on the
latest add-on value displayed before.
Activate function:
Repeatedly press the [CAL] key until "FUnC.SEL" appears.
Press [TARE] key
Repeatedly press the [CAL] key until the current "Addon:**" set-
ting appears.
(Example)
Press [TARE] key
You can use the [CAL] key to select between the following set-
tings:
" Adon-on"
" Adon-oF"
The current setting is indicated by the standstill display ( ).
60
Function activated
Function deactivated
ABT-NM-BA-e-1622