15.3 Remote control instructions
Attention:
Inputting characters and command codes not shown here into the balance may not
only alter the previous settings but may also impair proper measurement.
If by mistake characters or commands not shown here are entered into the balance,
immediately unplug the power supply cable and wait about ten seconds before plug-
ging again.
Command
Code
D01
Continuous output
D05
D06
1 time output with stability
D07
D08
1 time output at stability
D09
Q
ON/OFF switching
T
TS
Taring after stability wait
Measurement span cali-
C18
Measurements in add-up
+
R
mg
PERCENT
Percent determination
PCS
CT
SDENCE
LDENCE
%
G
- g
- mg
Deleting percentage de-
- PERCENT
- PCS
Piece counting removal
- CT
Solid specific gravity re-
- SDENCE
Liquid specific gravity re-
- LDENCE
ABT-NM-BA-e-1622
Function
1 time output
Auto Print
information
Halt output
Taring
bration
mode
Total reset
mg unit
Parts counting
ct unit
Density of solids
Density of liquids
100% setting
g, % switching
g unit removal
mg unit removal
termination t
ct unit removal
moval
moval
Description
The balance continuously outputs every
110ms .
Corresponds to [PRINT] key
See 14.4
The status of the stability mark is appended to
the head of the data with output.
S: when the standstill display appears
U: when standstill display does not appear
After command input, the data are output at
the first appearance of the standstill display.
Auto Print and continuous output halted
Switches between standby status and meas-
urement status.
Corresponds to the [O/T] key
After command input, taring is done at the
first appearance of the stability mark.
See ch. 14.5
All application measurements terminated and
reset
65